+To be moved to User's guide / WIKI ?
+
+
+-->'Media Storage SOP Class UID' (0x0002,0x0002)
+-->'SOP Class UID' (0x0008,0x0016) are set to
+ [Secondary Capture Image Storage]
+ (Potentialy, the image was modified by user, and post-processed;
+ it's no longer a 'native' image)
+
+--> 'Image Type' (0x0008,0x0008)
+ is forced to "DERIVED\PRIMARY"
+ (The written image is no longer an 'ORIGINAL' one)
+
+--> 'Modality' (0x0008,0x0060)
+ is defaulted to "OT" (other) if missing.
+ (a fully user created image belongs to *no* modality)
+
+--> 'Media Storage SOP Instance UID' (0x0002,0x0003)
+--> 'Implementation Class UID' (0x0002,0x0012)
+ are automatically generated; no user intervention possible
+
+--> 'Serie Instance UID'(0x0020,0x000e)
+--> 'Study Instance UID'(0x0020,0x000d) are kept as is if already exist
+ created if it doesn't.
+ The user is allowed to create his own Series/Studies,
+ keeping the same 'Serie Instance UID' / 'Study Instance UID'
+ for various images
+ Warning :
+ The user shouldn't add any image to a 'Manufacturer Serie'
+ but there is no way no to allowed him to do that
+
+
+--> If 'SOP Class UID' exists in the native image ('true DICOM' image)
+ we create the 'Source Image Sequence' SeqEntry (0x0008, 0x2112)
+
+ --> 'Referenced SOP Class UID' (0x0008, 0x1150)
+ whose value is the original 'SOP Class UID'
+ ---> 'Referenced SOP Instance UID' (0x0008, 0x1155)
+ whose value is the original 'SOP Class UID'
+
+--> Bits Stored, Bits Allocated, Hight Bit Position are checked for consistency
+--> Pixel Spacing (0x0028,0x0030) is defaulted to "1.0\1.0"
+--> Samples Per Pixel (0x0028,0x0002) is defaulted to 1 (grayscale)
+
+--> Instance Creation Date, Instance Creation Time, Study Date, Study Time
+ are force to current Date and Time
+
+--> Conversion Type (0x0008,0x0064)
+ is forced to 'SYN' (Synthetic Image)
+
+--> Study ID, Series Number, Instance Number, Patient Orientation (Type 2)
+ are created, with empty value if there are missing.
+
+--> Manufacturer, Institution Name, Patient's Name, (Type 2)
+ are defaulted with a 'gdcm' value.
+
+--> Patient ID, Patient's Birth Date, Patient's Sex, (Type 2)
+--> Referring Physician's Name (Type 2)
+ are created, with empty value if there are missing.
+
+ -------------------------------------------------------------------------------------*/